IPEC Water Cooled Heat Transfer Cart w/ Honeywell Chart Recorder USED (6976)
In great cosmetic condition. Turns on, buttons and digital screen respond well to input. Pump turns on. Did not test further. See pictures for more detail.

 

  • Manufacturer: IPEC
  • Model #: 4600016370
  • Model:  Heat Transfer Cart
  • Weight: 585 lbs
  • Includes: 
  • User Manual, 
  • Spare Parts List, 
  • Drawings on Flash Drive
  • • Chilled water supply valve (HV-2127-03).
  • • Chilled water control valve (TCV-2127-01).
  • • Heat exchanger (PHE-2127-01).
  • • WFI supply pump (CP-2127-01).
  • • Recirculation valve (HV-2127-01).
  • • Drain valve (HV-2127-02).
  • • Back pressure valve (HV-2127-04)
  • • WFI surge tank (TK-2127-02).
  • • Surge tank temperature indicator (TI-2127-01).
  • • Temperature recorder/controller (TRC-2127-01).
    • The portable heat transfer cart will be designed to cool the fixed process vessels from a starting temperature of 100°C (212°F) to an ending temperature of 15°C (59°F) in approximately 1 to 1.5 hour time period. The centrifugal pump mounted on the skid will cycle WFI through a plate heat exchanger used for cooling, which will then be transferred through the jacket of said process vessels.
    • During the initial WFI cool down the portable heat transfer cart will be manually hooked up and locally recirculated to cool down the WFI from 82°C (180°F) to 12°C (54°F). This is done via customer supplied flex hoses, manually overriding the chilled water control valve via through the Chart Recorder and starting the WFI recirculation pump.  Once the WFI has been initially cool down to 12°C (54°F), the customer supplied flex hoses will then be manually connected to the process vessel and the chilled water control valve manual override is set back to auto. At this time the temperature sensor input cable from the chart recorder/RTD transmitter is to be hooked up. Ensure that all the connections are made and secured before starting the WFI recirculation pump. The sequences will be manually controlled via the switchgear located on the system's control panel. The operator will initiate the cycle by manually switching on the pump through the pump on/off selector switch. Product cool down time from 100°C (212°F) to 15°C (59°F) is approximately 1 to 1.5 hour. After the process is finished the portable heat transfer cart can then be drained via manually opening the WFI surge tank drain valve.

    SVF Flow Company

    • Model #: E200W4SC
    • Volts: 115 vac
    • Amps: 5LP
    • Torque: 200 lbs
    • Cycle Time: 5.0 Sec
    • Duty Cycle: 2.5
    • NEMA: 4 

    Motor

    • Manufacturer: Franklin Electric
    • Model: 1103017430
    • Hz: 60
    • HP: 3/4
    • RPM: 3450
    • V: 115/230
    • FLA: 9.0/4.5
    • SF: 1.5
    • SFA: 12.2/6.1

    Heat Exchanger

    • Manufacturer: MaxChanger
    • Model: MX-10-0424HP-020
    • Height: 2.500"
    • Weight: 40 lbs
    • Standard Fitting Size: 1.00" 

      Vessel

    • Manufacture: Integrated Process Engineers & Constructors (IPEC)
    •  Tag #: TK-2127-01
    • Operating Volume: 10 Gal
    • Total Volume: 12 Gal
    • Vessel: Atmospheric-Non Code
    • Manufactured in the year 2004

    Circular Chart Recorder

    • Manufacturer: Honeywell
    • Model #: DR4500

    Acrylic Tube Flowmeter

    • Model #: 7200 Series
    • Metering Tube Machined cast acrylic
    • Internal Components  316L SS
    • Fitting Material  PVC (for water service only)
    • 316 SS, brass, aluminum (for water and air service)
    • Inlet/Outlet Fittings  NPT, vertical
    • O-Ring  Standard: EPR
    • PERFORMANCE
    • Capacities Water: 1 to 200 GPM
    • Air 4 to 245 SCFM
    • Scale 127 mm (5”) direct reading
    • Accuracy ± 3% to ± 6% of full scale flow
    • Turndown 10:1 unless otherwise indicated
    • Repeatability: 1% to 2%
    • Maximum Temperature: Water 54° C, Air 38° C
    • Maximum Pressure: Water 150 psig, Air 100 psig
    • Ambient Temperature: 1° C to 52° C

    Shipping

    • Weight: 800 lbs
    • Dimensions: 62 x 44 x 78"

    ​EQUIPMENT OVERVIEW
    The equipment employed in this installation is utilized for heat transfer, specifically the cooling of process vessels. A description of the major process equipment contained in the scope of this installation is presented in this section.
    Portable Heat Transfer Cart Components
    • Chilled water supply valve (HV-2127-03).
    • Chilled water control valve (TCV-2127-01).
    • Heat exchanger (PHE-2127-01).
    • WFI supply pump (CP-2127-01).
    • Recirculation valve (HV-2127-01).
    • Drain valve (HV-2127-02).
    • Back pressure valve (HV-2127-04)
    • WFI surge tank (TK-2127-02).
    • Surge tank temperature indicator (TI-2127-01).
    • Temperature recorder/controller (TRC-2127-01).
    PROCESS OVERVIEW
    The portable heat transfer cart will be designed to cool the fixed process vessels from a starting
    temperature of 100ºC (212ºF) to an ending temperature of 15ºC (59ºF) in approximately 1 to 1.5
    hour time period. The centrifugal pump mounted on the skid will cycle WFI through a plate heat
    exchanger used for cooling, which will then be transferred through the jacket of said process
    vessels
    OPERATION OVERVIEW
    During the initial WFI cool down the portable heat transfer cart will be manually hooked up and locally recirculated to cool down the WFI from 82ºC (180ºF) to 12ºC (54ºF). This is done via customer supplied flex hoses, manually overriding the chilled water control valve via through the Chart Recorder and starting the WFI recirculation pump. Once the WFI has been initially cool down to 12ºC (54ºF), the customer supplied flex hoses will then be manually connected to the process vessel and the chilled water control valve manual override is set back to auto. At this time the temperature sensor input cable from the chart recorder/RTD transmitter is to be hooked up. Ensure that all the connections are made and secured before starting the WFI recirculation pump. The sequences will be manually controlled via the switchgear located on the system's control panel. The operator will initiate
    the cycle by manually switching on the pump through the pump on/off selector switch. Product cool down time from 100ºC (212ºF) to 15ºC (59ºF) is approximately 1 to 1.5 hour. After the process is finished the portable heat transfer cart can then be drained via manually opening the WFI surge tank drain valve.




    Crating Fee: $250
    Sold as is with 30 day Right of Return, only pictured items included
    Easy return/refund for any reason
    Anacapa Equipment

    RETURNS 
    Anacapa Equipment Brokers, Inc. (“Seller”) strives to offer excellent customer service.   You may return any equipment purchased from us for any reason within thirty (30) days after your receipt of the equipment for a full refund of the purchase price, provided that returned items are received in the same condition as sent.  We DO NOT ship internationally.  If you are using a freight forwarder then returns are VOID.  

    SHIPPING                                                                                                               All equipment purchased is shipped within one (1) business day (subject to freight availability) after receipt of payment and is shipped FOB place of shipment.  We only charge actual shipping cost.  If eBay overcharges for shipping, we issue an immediate refund for the difference. If you are undercharged, we will absorb the cost. We DO NOT ship internationally.  IF you are using a freight forwarder then returns are VOID.  
    LIMITED WARRANTY                                                                 
    Buyer acknowledges that the equipment is being sold “as is" and without any warranty of merchantability or fitness for a particular use.  The entire risk as to the quality and performance of the equipment is with the Buyer.  Should the equipment prove defective following its purchase the Buyer and not the Seller, manufacturer, distributor, or retailer assumes the entire cost of all necessary servicing or repair.  Buyer may inspect the equipment prior to purchase and placement of the order.  Failure to inspect prior to placement of an order is deemed a waiver of Buyer’s right to inspect.  Buyer is responsible for determining the suitability of the equipment for its application.  Seller assumes no liability for the use or misuse of any equipment purchased by Buyer. It is assumed that the equipment will be installed and operated by qualified personnel. The Buyer agrees that the Seller shall not be held responsible or liable for any injuries or damages, whether incidental or consequential, associated in any way with the equipment purchased.  The Buyer, by purchasing this equipment, indicates its acknowledgment of, and agreement, to the terms of this limited warranty.

    INDEMNITY                                                                                                            
    Buyer shall be solely liable for, and shall defend, indemnify, and hold Seller, its Affiliates, and all of their respective directors, officers, employees, agents, and independent contractors (“Indemnitees”) harmless from and against any and all damages, liabilities, injuries, losses, costs (including reasonable attorneys and experts fees at trial and on appeal), and expenses which may be incurred by, asserted against, or recoverable from any Indemnitee as a result of any actual or threatened action, suit or proceeding arising out of or relating in any way to the purchase of equipment from Seller.  The validity and interpretation of this Agreement and the enforcement thereof shall be governed by the laws of the State of California without regard to any principles governing conflicts of laws.  Any dispute arising from the purchase of equipment from Seller must be resolved in the State Courts of California. 

     

     
     









     

     

    IPEC Water Cooled Heat Transfer Cart w/ Honeywell Chart Recorder USED (6976)
    In great cosmetic condition. Turns on, buttons and digital screen respond well to input. Pump turns on. Did not test further. See pictures for more detail.

     

    • Manufacturer: IPEC
    • Model #: 4600016370
    • Model:  Heat Transfer Cart
    • Weight: 585 lbs
    • Includes: 
  • User Manual, 
  • Spare Parts List, 
  • Drawings on Flash Drive
  • • Chilled water supply valve (HV-2127-03).
  • • Chilled water control valve (TCV-2127-01).
  • • Heat exchanger (PHE-2127-01).
  • • WFI supply pump (CP-2127-01).
  • • Recirculation valve (HV-2127-01).
  • • Drain valve (HV-2127-02).
  • • Back pressure valve (HV-2127-04)
  • • WFI surge tank (TK-2127-02).
  • • Surge tank temperature indicator (TI-2127-01).
  • • Temperature recorder/controller (TRC-2127-01).
    • The portable heat transfer cart will be designed to cool the fixed process vessels from a starting temperature of 100°C (212°F) to an ending temperature of 15°C (59°F) in approximately 1 to 1.5 hour time period. The centrifugal pump mounted on the skid will cycle WFI through a plate heat exchanger used for cooling, which will then be transferred through the jacket of said process vessels.
    • During the initial WFI cool down the portable heat transfer cart will be manually hooked up and locally recirculated to cool down the WFI from 82°C (180°F) to 12°C (54°F). This is done via customer supplied flex hoses, manually overriding the chilled water control valve via through the Chart Recorder and starting the WFI recirculation pump.  Once the WFI has been initially cool down to 12°C (54°F), the customer supplied flex hoses will then be manually connected to the process vessel and the chilled water control valve manual override is set back to auto. At this time the temperature sensor input cable from the chart recorder/RTD transmitter is to be hooked up. Ensure that all the connections are made and secured before starting the WFI recirculation pump. The sequences will be manually controlled via the switchgear located on the system's control panel. The operator will initiate the cycle by manually switching on the pump through the pump on/off selector switch. Product cool down time from 100°C (212°F) to 15°C (59°F) is approximately 1 to 1.5 hour. After the process is finished the portable heat transfer cart can then be drained via manually opening the WFI surge tank drain valve.

    SVF Flow Company

    • Model #: E200W4SC
    • Volts: 115 vac
    • Amps: 5LP
    • Torque: 200 lbs
    • Cycle Time: 5.0 Sec
    • Duty Cycle: 2.5
    • NEMA: 4 

    Motor

    • Manufacturer: Franklin Electric
    • Model: 1103017430
    • Hz: 60
    • HP: 3/4
    • RPM: 3450
    • V: 115/230
    • FLA: 9.0/4.5
    • SF: 1.5
    • SFA: 12.2/6.1

    Heat Exchanger

    • Manufacturer: MaxChanger
    • Model: MX-10-0424HP-020
    • Height: 2.500"
    • Weight: 40 lbs
    • Standard Fitting Size: 1.00" 

      Vessel

    • Manufacture: Integrated Process Engineers & Constructors (IPEC)
    •  Tag #: TK-2127-01
    • Operating Volume: 10 Gal
    • Total Volume: 12 Gal
    • Vessel: Atmospheric-Non Code
    • Manufactured in the year 2004

    Circular Chart Recorder

    • Manufacturer: Honeywell
    • Model #: DR4500

    Acrylic Tube Flowmeter

    • Model #: 7200 Series
    • Metering Tube Machined cast acrylic
    • Internal Components  316L SS
    • Fitting Material  PVC (for water service only)
    • 316 SS, brass, aluminum (for water and air service)
    • Inlet/Outlet Fittings  NPT, vertical
    • O-Ring  Standard: EPR
    • PERFORMANCE
    • Capacities Water: 1 to 200 GPM
    • Air 4 to 245 SCFM
    • Scale 127 mm (5”) direct reading
    • Accuracy ± 3% to ± 6% of full scale flow
    • Turndown 10:1 unless otherwise indicated
    • Repeatability: 1% to 2%
    • Maximum Temperature: Water 54° C, Air 38° C
    • Maximum Pressure: Water 150 psig, Air 100 psig
    • Ambient Temperature: 1° C to 52° C

    Shipping

    • Weight: 800 lbs
    • Dimensions: 62 x 44 x 78"

    ​EQUIPMENT OVERVIEW
    The equipment employed in this installation is utilized for heat transfer, specifically the cooling of process vessels. A description of the major process equipment contained in the scope of this installation is presented in this section.
    Portable Heat Transfer Cart Components
    • Chilled water supply valve (HV-2127-03).
    • Chilled water control valve (TCV-2127-01).
    • Heat exchanger (PHE-2127-01).
    • WFI supply pump (CP-2127-01).
    • Recirculation valve (HV-2127-01).
    • Drain valve (HV-2127-02).
    • Back pressure valve (HV-2127-04)
    • WFI surge tank (TK-2127-02).
    • Surge tank temperature indicator (TI-2127-01).
    • Temperature recorder/controller (TRC-2127-01).
    PROCESS OVERVIEW
    The portable heat transfer cart will be designed to cool the fixed process vessels from a starting
    temperature of 100ºC (212ºF) to an ending temperature of 15ºC (59ºF) in approximately 1 to 1.5
    hour time period. The centrifugal pump mounted on the skid will cycle WFI through a plate heat
    exchanger used for cooling, which will then be transferred through the jacket of said process
    vessels
    OPERATION OVERVIEW
    During the initial WFI cool down the portable heat transfer cart will be manually hooked up and locally recirculated to cool down the WFI from 82ºC (180ºF) to 12ºC (54ºF). This is done via customer supplied flex hoses, manually overriding the chilled water control valve via through the Chart Recorder and starting the WFI recirculation pump. Once the WFI has been initially cool down to 12ºC (54ºF), the customer supplied flex hoses will then be manually connected to the process vessel and the chilled water control valve manual override is set back to auto. At this time the temperature sensor input cable from the chart recorder/RTD transmitter is to be hooked up. Ensure that all the connections are made and secured before starting the WFI recirculation pump. The sequences will be manually controlled via the switchgear located on the system's control panel. The operator will initiate
    the cycle by manually switching on the pump through the pump on/off selector switch. Product cool down time from 100ºC (212ºF) to 15ºC (59ºF) is approximately 1 to 1.5 hour. After the process is finished the portable heat transfer cart can then be drained via manually opening the WFI surge tank drain valve.




    Crating Fee: $250
    Sold as is with 30 day Right of Return, only pictured items included
    Easy return/refund for any reason
    Anacapa Equipment

    RETURNS 
    Anacapa Equipment Brokers, Inc. (“Seller”) strives to offer excellent customer service.   You may return any equipment purchased from us for any reason within thirty (30) days after your receipt of the equipment for a full refund of the purchase price, provided that returned items are received in the same condition as sent.  We DO NOT ship internationally.  If you are using a freight forwarder then returns are VOID.  

    SHIPPING                                                                                                               All equipment purchased is shipped within one (1) business day (subject to freight availability) after receipt of payment and is shipped FOB place of shipment.  We only charge actual shipping cost.  If eBay overcharges for shipping, we issue an immediate refund for the difference. If you are undercharged, we will absorb the cost. We DO NOT ship internationally.  IF you are using a freight forwarder then returns are VOID.  
    LIMITED WARRANTY                                                                 
    Buyer acknowledges that the equipment is being sold “as is" and without any warranty of merchantability or fitness for a particular use.  The entire risk as to the quality and performance of the equipment is with the Buyer.  Should the equipment prove defective following its purchase the Buyer and not the Seller, manufacturer, distributor, or retailer assumes the entire cost of all necessary servicing or repair.  Buyer may inspect the equipment prior to purchase and placement of the order.  Failure to inspect prior to placement of an order is deemed a waiver of Buyer’s right to inspect.  Buyer is responsible for determining the suitability of the equipment for its application.  Seller assumes no liability for the use or misuse of any equipment purchased by Buyer. It is assumed that the equipment will be installed and operated by qualified personnel. The Buyer agrees that the Seller shall not be held responsible or liable for any injuries or damages, whether incidental or consequential, associated in any way with the equipment purchased.  The Buyer, by purchasing this equipment, indicates its acknowledgment of, and agreement, to the terms of this limited warranty.

    INDEMNITY                                                                                                            
    Buyer shall be solely liable for, and shall defend, indemnify, and hold Seller, its Affiliates, and all of their respective directors, officers, employees, agents, and independent contractors (“Indemnitees”) harmless from and against any and all damages, liabilities, injuries, losses, costs (including reasonable attorneys and experts fees at trial and on appeal), and expenses which may be incurred by, asserted against, or recoverable from any Indemnitee as a result of any actual or threatened action, suit or proceeding arising out of or relating in any way to the purchase of equipment from Seller.  The validity and interpretation of this Agreement and the enforcement thereof shall be governed by the laws of the State of California without regard to any principles governing conflicts of laws.  Any dispute arising from the purchase of equipment from Seller must be resolved in the State Courts of California. 

     

     
     

    Specifications

    ManufacturerIPEC
    Model4600016370 E200W4SC 1103017430 MX-10-0424HP-20
    ConditionUsed
    Stock Number6976
    TypeCompact Heat Exchanger
    System configurationVertical
    Year manufactured2004
    FinishBrass